| Temperature Compensation | automatic compensation (Pt1000/NTC30K); manual compensation or compensation disabled |
|---|---|
| Reference Temperature | Tref25 °C |
| Dimensions | 96 × 96 × 148(H × W × D) |
| Output Signal | 0/4-20mA; RS485 |
| Power Supply | 100V–260VAC, 50/60Hz; 16V–30VDC |
| Ambient Temperature | 0–50 °C |
| Relative Humidity | 5–85% |
| Conductivity Measurement | range 20.00 μS/200.0 μS/2000 μS/20.00 mS/200.0 mS; five manually or automatically selected ranges; resolution 0.01 μS/cm; accuracy ±1% (±1 digit) |
| Resistivity Measurement | range 0.00–20.00MΩ/cm; resolution 0.01MΩ/cm; accuracy ±1% (±1 digit) |
| Salinity Measurement | range 0.0–70.0 ppt; resolution 0.1 ppt; accuracy ±1% (±1 digit) |
| Temperature Measurement | range -30.0–+130.0 °C; resolution 0.1 °C; accuracy ±0.2% (±1 digit) |
| Temperature Coefficient | linear temperature compensation (0.00%–20.00%) and nonlinear compensation; no compensation |
| Display | adjustable backlit dot-matrix LCD |
| IP Rating | IP65 with rear cover installed |
The Conductivity measurement specifications are: range 20.00 μS/200.0 μS/2000 μS/20.00 mS/200.0 mS; five manually or automatically selected ranges; resolution 0.01 μS/cm; accuracy ±1% (±1 digit).
The Resistivity measurement specifications are: range 0.00–20.00MΩ/cm; resolution 0.01MΩ/cm; accuracy ±1% (±1 digit).
The Salinity measurement specifications are: range 0.0–70.0 ppt; resolution 0.1 ppt; accuracy ±1% (±1 digit).